Product description:
1. UST-series with the function of cooling down from high temperature and the temperature range from 250℃ to 50℃.
2. Equipped with heating and cooling system, large heat exchange area, rapid speed of heating-cooling, the small demand for heat conducting medium.
3. Can heating and cooling continuously.
4. The whole system is closed, there is no oil mist at high temperature, no volatilization of heat conducting medium at high temperature.
5. With inter temperature sensor PT-100 to adjust internal circulation temperature.
6. The whole system is a closed liquid circulation system with expansion tank, the expansion tank, and liquid circulation are thermal insulation, do not participate in liquid circulation, regardless of high or low temperature, the medium in the expansion tank is limited to 60℃.
7. Allow fill in 15℃ cooling water, to enlarge temperature range of 25℃ to 250℃.
Heating temperature control system consistent with SUNDI-series, but with only heating function, cooling can flow in water to cool to room temperature.
Technical parameter
The system with feedforward with PV, the master PID loop operation results of the PV output and feedforward signals as the composite of the settings from the control loop, through such control of the temperature gradient to ensure the accuracy of temperature control system.
type
|
Temp range℃
|
Heating KW
|
Cooling capacity KW AT ℃
|
Flow/pressure
L/min bar |
Outside size
|
power
KW | ||||
300
|
200
|
100
|
65
| |||||||
UST-3530
|
+50~+300
|
3.5
|
3.5
|
3.5
|
2.8
|
1
|
35
|
2
|
400*600*1150
|
5kw
|
UST-5530
|
+50~+300
|
5.5
|
5.5
|
5.5
|
4
|
1.8
|
50
|
2
|
500*680*1250
|
7kw
|
UST-7530
|
+50~+300
|
7.5
|
7.5
|
7.5
|
5
|
2.2
|
75
|
2
|
550*680*1450
|
9kw
|
UST-A1030
|
+50~+300
|
10
|
10
|
10
|
7
|
3
|
110
|
2.5
|
550*680*1450
|
12kw
|
UST-A1530
|
+50~+300
|
15
|
15
|
15
|
12
|
5
|
110
|
2.5
|
650*700*1750
|
17kw
|
UST-A2530
|
+50~+300
|
25
|
25
|
25
|
20
|
8.5
|
150
|
2.5
|
750*800*1850
|
28kw
|
UST-A3830
|
+50~+300
|
38
|
38
|
38
|
30
|
12
|
200
|
2.5
|
1000*950*1750
|
41kw
|
UST-A6030
|
+50~+300
|
60
|
60
|
60
|
48
|
19
|
250
|
2.5
|
1500*1250*2050
|
65kw
|
UST-A9530
|
+50~+300
|
95
|
95
|
95
|
76
|
30
|
400
|
2.5
|
2050*1450*2050
|
101kw
|
UST-A13030
|
+50~+300
|
130
|
130
|
130
|
100
|
40
|
400
|
2.5
|
2050*1450*2050
|
136kw
|
UST-A20030
|
+50~+300
|
200
|
200
|
200
|
160
|
64
|
600
|
3
|
2750*1450*2250
|
212kw
|
Type
|
SUNDI/UST series
|
Control Mode
|
Feedback PID + Our special dynamic control calculation, Lneya PLC controller
|
Temp.control
|
Material reaction temp.control mode and jacket temp.control mode
|
Program Editor
|
10 programs, each program can edit 45 steps.
|
Communication Portocol
|
MODBUS RTU Protocol,RS485 interface
Optional Ethernet interface/R232 interface
|
Temp.feedback
|
The temperature of three points: the inlet and outlet of heat-conducting medium, material in reactor(External temperature sensor), PT100 sensor
|
Material temp. feedback
|
Material temp. feedback:PT100 or 4~20mA or communication given
Temp.feedback:PT100
|
Temp. difference
|
Set or control oil outlet temp and reactor material temp. difference
Set or control equipment liquid inlet temp. And liquid outlet temp. Difference( keep system safe)
|
Material temp.accuracy
|
±1℃ (sometimes ±0.5℃)
|
Thermal medium temp. accuracy
|
±0.5℃
|
Closed circulation system
|
The whole system is full closed circulation, there is no oil mist at high temperature and no water vapor at low temperature, pressure do not rise up when system is running. The system will supply oil automatically at low temperature.
|
heating
|
Heating max output power
The heater has triple protection and independent temperature limiter to ensure the safety of heating system.
The power is greater than 15KW with a voltage regulator, and the heating power output control adopts 4-20ma linear control.
|
Cooling capacity
|
The power of taking away the heat at different temperature
|
Circulation flow rate and pressure
|
Lower than 75L/min, use Lneya Magnetic pump
Higher than 75L/min use Ultra wide temperature high energy efficiency centrifugal pump.
Pressure selection: 2bar 3bar 5bar 8bar(small power equipment can not customize)
The greater the pressure, the greater the power of the motor, which will increase the system load and reduce the cooling capacity of the system.
|
compressor
|
Different type different brand compress
Mitsubishi, skop, Emerson, Danfoss, Italy Dorin, German blog, bizell, French Tecumseh.
|
evaporator
|
KAORI Plate heat exchanger
|
Oil separator
|
Emerson/Lneya
|
Drying filter
|
danfoss/Emerson
|
Shock tube
|
CASTEL/packless
|
voltage-controlled
|
danfoss/Emerson/fengsheng
|
Expansion valve
|
Emerson thermal expansion valve/Emerson electronic expansion valve/danfoss thermal expansion valve.
Select different expansion valves with different temperature and cooling capacity.
|
Operation panel
|
Lneya 7-inch touch screen,show temp curve\EXCEL data output
Optional Siemens explosion - proof touch screen.(Place the operation panel outside the site area.)
optional GMP, FDA operation and record module (operator use record, data record, alarm record, etc.)
|
Safety protection
|
Self-diagnosis function; freezer overload protection; high pressure switch, overload relay,thermal protection device, liquid low level protection, high temperature protection and temperature fault protection.
|
refrigerant
|
R-404A / R404A R23 mixture/ R404A R23 R14 mixture
|
interface size
|
G1/2、G3/4、G1、DN25 DN32 DN40 DN50 DN65 DN80 DN100(PN10 RF)
|
Water-cooling W
|
With W is water cooled type, lower than -60 degree equipment, you need use 7-25 degree cooled water
|
Water cooled condenser.
|
Tube heat exchanger, plate heat exchanger, tube condenser.
packless、sheng、KAORI、JINDIAN、XIAOLING etc
|
power
|
AC 380V 50HZ Three-phase five-wire system. Small power 220V 50HZ single phase
Optional 460V 60HZ、220V 60HZ three-phase
|
Shell material
|
Standard shellSUS 304,optionalspraying plastics (color 7035)
|
Isolation of explosion-proof
|
optioanl(EXdIIBT4)
Lneya explosion-proof product installation license number.:PCEC-2017-B025
|
Positive pressure explosion-proof
|
optioanl(EXPXdmbIICT4) The positive pressure system must be water-cooled type
Lneya explosion-proof product installation license number:PCEC-2017-B025
|
Self pressurizing system
|
Optional self-pressurized 3bar system, the temperature of the system can be expanded, the expansion tank is normal pressure.
KDOC-80 can be used in a pressurized system with a maximum temperature of 250 degrees (-85 degree ~ 250 degree).
KDOC-220 can be used in the self-pressurized system with maximum temperature to 320 degrees (-50 degree ~ 320 degree).
KDOC-210 can be used in a self-pressurized system with a maximum temperature of 300 degrees (-70 degree ~ 300 degree)
|
Flow/pressure
|
Optional flow/pressure sensor, display flow and pressure on the touch screen.
Optional control flow/pressure.
|
One for two system
|
Optional one for two systems: the costis lower than one to one system.
One unit heating and cooling machine contains two circulation system, two cooling system, two heating systems, one expansion system, one electrical system.
|
High exothermic control
|
Customizable high heat control system algorithm and components.
|
flow reactor temperature control module.
|
It can be used in the flow reactor temp control system (pump, internal circulation system, control algorithm, refrigeration)
|
The outdoor installation of air-cooled chillers makes it easy to be affected by outdoor sunlight, and the ventilation and the environment and space has a direct impact on the cooling effect of the chiller. Today we are going to talk about the common failure and processing methods:
(1) Frequent start and stop of the compressor: there may be too much or too little refrigerant, resulting in pressure protection switch action which is resulted from too high exhaust pressure or too low suction pressure, and also poor water system circulation make sevaporator surface frost.
Method: When the refrigerant is too much, the extra can be released at the vent. If there is a shortage, check whether there are leaks of the fluorine and fill the refrigerant.
If the water circulation system is blocked or stored in the air, you should exhaust; if water circulation too small, utilize the open refrigeration water circulation system.
(2) Noisy compressor when running: check whether the expansion valve fails, the temperature package discharges the suction pipe from the compressor, which causes the liquid refrigerant return to the compressor. If it is damaged, it should be repaired, for example, if the compressor is damaged, it should be replaced.
(3) Poor refrigeration capacity: the cause of the poor refrigeration capacity may be the insufficient refrigerantand low evaporation temperature that makes the por cooling effect of pipeline; poor heat dissipation of the condenser; too large the resistant of the water circulation system,etc..
The refrigerant pipe should be checked for leakage and fill the refrigerant after repairing.You should repair the insulation layer and Clean the condenser to check whether the fan is working properly as well as clean the pipe filter.
(1) Frequent start and stop of the compressor: there may be too much or too little refrigerant, resulting in pressure protection switch action which is resulted from too high exhaust pressure or too low suction pressure, and also poor water system circulation make sevaporator surface frost.
Method: When the refrigerant is too much, the extra can be released at the vent. If there is a shortage, check whether there are leaks of the fluorine and fill the refrigerant.
If the water circulation system is blocked or stored in the air, you should exhaust; if water circulation too small, utilize the open refrigeration water circulation system.
(2) Noisy compressor when running: check whether the expansion valve fails, the temperature package discharges the suction pipe from the compressor, which causes the liquid refrigerant return to the compressor. If it is damaged, it should be repaired, for example, if the compressor is damaged, it should be replaced.
(3) Poor refrigeration capacity: the cause of the poor refrigeration capacity may be the insufficient refrigerantand low evaporation temperature that makes the por cooling effect of pipeline; poor heat dissipation of the condenser; too large the resistant of the water circulation system,etc..
The refrigerant pipe should be checked for leakage and fill the refrigerant after repairing.You should repair the insulation layer and Clean the condenser to check whether the fan is working properly as well as clean the pipe filter.
The outdoor installation of air-cooled chillers makes it easy to be affected by outdoor sunlight, and the ventilation and the environment and space has a direct impact on the cooling effect of the chiller. Today we are going to talk about the common failure and processing methods:
(1) Frequent start and stop of the compressor: there may be too much or too little refrigerant, resulting in pressure protection switch action which is resulted from too high exhaust pressure or too low suction pressure, and also poor water system circulation make sevaporator surface frost.
Method: When the refrigerant is too much, the extra can be released at the vent. If there is a shortage, check whether there are leaks of the fluorine and fill the refrigerant.
If the water circulation system is blocked or stored in the air, you should exhaust; if water circulation too small, utilize the open refrigeration water circulation system.
(2) Noisy compressor when running: check whether the expansion valve fails, the temperature package discharges the suction pipe from the compressor, which causes the liquid refrigerant return to the compressor. If it is damaged, it should be repaired, for example, if the compressor is damaged, it should be replaced.
(3) Poor refrigeration capacity: the cause of the poor refrigeration capacity may be the insufficient refrigerantand low evaporation temperature that makes the por cooling effect of pipeline; poor heat dissipation of the condenser; too large the resistant of the water circulation system,etc..
The refrigerant pipe should be checked for leakage and fill the refrigerant after repairing.You should repair the insulation layer and Clean the condenser to check whether the fan is working properly as well as clean the pipe filter.
(1) Frequent start and stop of the compressor: there may be too much or too little refrigerant, resulting in pressure protection switch action which is resulted from too high exhaust pressure or too low suction pressure, and also poor water system circulation make sevaporator surface frost.
Method: When the refrigerant is too much, the extra can be released at the vent. If there is a shortage, check whether there are leaks of the fluorine and fill the refrigerant.
If the water circulation system is blocked or stored in the air, you should exhaust; if water circulation too small, utilize the open refrigeration water circulation system.
(2) Noisy compressor when running: check whether the expansion valve fails, the temperature package discharges the suction pipe from the compressor, which causes the liquid refrigerant return to the compressor. If it is damaged, it should be repaired, for example, if the compressor is damaged, it should be replaced.
(3) Poor refrigeration capacity: the cause of the poor refrigeration capacity may be the insufficient refrigerantand low evaporation temperature that makes the por cooling effect of pipeline; poor heat dissipation of the condenser; too large the resistant of the water circulation system,etc..
The refrigerant pipe should be checked for leakage and fill the refrigerant after repairing.You should repair the insulation layer and Clean the condenser to check whether the fan is working properly as well as clean the pipe filter.